EMC China Lab

Disinfection Validation for Reusable Medical Devices

Views :
Update time : 2025-04-15

Definition of Disinfection

01 Low-Level Disinfection

A lethal process that can kill most vegetative bacteria, some fungi, and viruses, but cannot kill mycobacteria and bacterial spores.

 

02 Intermediate-Level Disinfection

A lethal process that can kill most viruses, mycobacteria, fungi, and vegetative bacteria, but cannot kill spores.

 

03 High-Level Disinfection

A lethal process that uses sterilants under conditions less severe than sterilization. This process can kill all MICrobial vegetative forms, except for a large number of spores.

 

Disinfection Validation for Reusable Medical Devices(图1)


Selection of Disinfection Methods

When selecting disinfection methods, the compatibility between the disinfection method and the medical device shoULd be consideRED, including the adaptability of the disinfection factors, their impact on device performance, and their physical or chemical effects on the device materials.

For thermal disinfection, factors such as deformation and thermoplastic effects should be taken into account. For chemical disinfection, issues such as chemical corrosion, material compatibility, and disinfectant residue should be considered.

 

Disinfection Validation

Purpose of Disinfection Validation:

To confirm whether the disinfection method and disinfectant specified in the instructions/label can effectively remove microorganisms from medical devices.

Each disinfection method and/or disinfectant listed in the device instructions must be individually validated.

Thermal disinfection and chemical disinfection are different disinfection methods and must be validated separately.

 

1) Thermal Disinfection: Each set of conditions specified in the instructions must be individually validated.

2) Chemical Disinfection: Each recommended disinfectant must be individually validated.

3) Manual and Automatic Disinfection: Must be validated separately.

 

Implementation of Disinfection Validation

01 Inoculation

 

1.1 Selection of Test Microorganisms

Based on factors such as the intended use of the device, risk level, disinfection level, resistance of the disinfection process, and the resistance of the bioburden on the device, the representativeness of the reference microorganisms should be confirmed.

AAMI TIR 12-2020 recommends using typical vegetative organisms such as Staphylococcus aureus, PSEudomonas aeruginosa, Escherichia coli, and representative Klebsiella strains (e.g., Klebsiella pneumoniae) for low-level disinfection validation; the same vegetative strains along with appropriate Mycobacterium strains for intermediate-level disinfection validation; and suitable Mycobacterium strains for high-level disinfection validation.

 

1.2 Selection of Inoculation Sites

Testing can be conducted on the entire device or on the most difficult-to-disinfect areas. These include: internal channels of the device, hinges, pits, joints with gaps, shafts within cavities, cracks, etc.

 

02 Disinfection

After microbial inoculation, disinfection of the product should be carried out using clearly defined routine and validation disinfection methods.

Both routine and validation disinfection methods must have clearly defined process parameters. Validation disinfection should use the lowest process parameters, such as:

- Shortest time

- Lowest temperature

- Lowest concentration

 

03 Elution

After disinfection, any remaining microorganisms on the device should be eluted. The elution efficiency must be verified to ensure the elution method can sufficiently remove residual microorganisms.

The eluent may require the addition of neutralizers to counteract residual disinfectants. It must also be ensured that the neutralizer does not inhibit microbial growth.

In addition, the elution method must be validated to confirm it does not damage the microorganisms.

 

04 Culturing

The eluted solution is collected and filtered using a membrane filtration method. The membrane is then placed onto the appropriate culture medium and incubated at a specified temperature for a defined duration.

 

05 Acceptance Criteria

According to AAMI TIR 12-2020:

- For low-level disinfection, typical vegetative organisms such as Staphylococcus aureus, Escherichia coli, Pseudomonas aeruginosa, and Klebsiella strains must show a reduction of no less than 6 log10.

- For intermediate-level disinfection, in addition to meeting the requirements for low-level disinfection, appropriate Mycobacterium strains must show a reduction of no less than 3 log10.

- For high-level disinfection, Mycobacterium strains must show a reduction of no less than 6 log10.

 

Currently, JJR Laboratory in China mainly provides the following testing services:

- Microbiological and chemical performance testing of medical devices

- Packaging system performance testing, including sterile barrier system performance, accelerated aging, natural aging, high and low temperature environmental testing, and transportation performance

- Water quality testing

- Cleaning, disinfection, and sterilization validation of reprocessed medical devices

- Routine sterilization, including ethylene oxide sterilization, steam sterilization, and low-temperature vaporized hydrogen peroxide plasma sterilization.

 

Testing methods cover a wide range of national and international standards such as GB, YY, ISO, AAMI, ASTM, ISTA, EN, USP, and EP.


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
What is the EN 61326-2-3 Standard? What is the EN 61326-2-3 Standard?
01 .31.2026
EN 61326-2-3 ensures EMC compliance for sensors, transmitters, lab & industrial equipment. JJR o...
Why Do Smart Sockets Need IEC 60884 Certification? Why Do Smart Sockets Need IEC 60884 Certification?
01 .31.2026
Smart sockets need IEC 60884 certification for safety & market ; JJR CNAS/ILAC/ISO17025 labs pro...
Why Retest the Device if the 5G Module Already Has Why Retest the Device if the 5G Module Already Has
01 .31.2026
Even if a 5G module is CE-certified, the whole device needs testing per EU RED; JJR LAB provides ISO...
Overview of IEC 62087 Test Standard Overview of IEC 62087 Test Standard
01 .31.2026
IEC 62087 defines power-consumption tests for TVs, monitors, audio/video devices. JJR Lab (CMS, CNAS...
CISPR 25 Test Standard Compliance Guide CISPR 25 Test Standard Compliance Guide
01 .31.2026
Automakers face EMI in smart vehicles; CISPR 25 & GB/T 18655 ensure CE/RE compliance. JJR Lab (C...
IEC/UL/CSA 62368-1 Electrical Distance Testing IEC/UL/CSA 62368-1 Electrical Distance Testing
01 .31.2026
IEC/UL/CSA 62368-1 electrical distance testing supports global market access. CMS, CNAS, ISO/IEC 170...
Canada Wireless Device IC Certification RSS-210 Te Canada Wireless Device IC Certification RSS-210 Te
01 .30.2026
Canada Wireless Device IC Certification RSS-210 testing by JJR, a CMS, CNAS, ISO/IEC 17025 lab, cove...
FCC Part 15.231 for Wireless Remote Controls and S FCC Part 15.231 for Wireless Remote Controls and S
01 .30.2026
FCC Part 15.231 testing for wireless remotes and sensors, covering RF, EMC, bandwidth and duty cycle...

Leave Your Message