EMC China Lab

FDA IEC TR 60601-4-2 Standard Testing

Views :
Update time : 2025-05-13

In the field of medical devices, the International Electrotechnical Commission (IEC) has developed two critical standards to ensure electromagnetic compatibility (EMC): iec 60601-4-2 and IEC 60601-1-2. While these two standards share certain similarities, they focus on different aspects of emc compliance.

 

FDA IEC TR 60601-4-2 Standard Testing(图1)


Currently, only the FDA has officially released a new EMC guidance document to replace the 2016 version. Neither the EU nor China has plans to develop an EN-equivalent standard of IEC TR 60601-4-2. Today, our focus is to understand the IEC TR 60601-4-2 standard.

 

Understanding IEC TR 60601-4-2

Recently, during 510(k) project discussions, many clients have asked why the IEC test submission list includes a report for IEC 60601-4-2. To address this, we will explore the similarities and differences between the two standards.

 

IEC TR 60601-4-2 is a technical report specifically focused on EMC immunity. Unlike IEC 60601-1-2, which concentrates on basic safety and essential performance, IEC 60601-4-2 emphasizes the performance criteria under electromagnetic immunity tests based on the intended use of medical electrical equipment.

 

FDA IEC TR 60601-4-2 Standard Testing(图2)


For example, a device may be affected by EMC but still remain safe; however, a device that claims to be safe might not be effective. IEC TR 60601-4-2 addresses performance criteria beyond basic pass/fail definitions of essential performance.

 

Published in 2016, this technical report complements IEC 60601-1-2 by guiding manufacturers and testing bodies on how to consider and define performance criteria and immunity test levels. However, since it's a supplementary document and not mandatory in most regions, it is often overlooked.

 

Structure of IEC TR 60601-4-2

By examining the table of contents, we can see that the structure is similar to IEC 60601-1, provided that performance criteria are clearly defined. The standard also clarifies that if performance under IEC 60601-1-2 immunity levels is met, further testing may not be necessary.

 

Depending on the intended use and environment, th

FDA IEC TR 60601-4-2 Standard Testing(图3)

e guideline recommends different considerations for EMC immunity. Annex C of IEC 60601-4-2 provides detailed examples, which we won’t expand on here.

 

FDA IEC TR 60601-4-2 Standard Testing(图4)


As stated in the preface, the FDA requires evidence of compliance with IEC TR 60601-4-2 because the newly published EMC guidance refers to the recommendations in this document mULtiple times, as shown below:

 

> — It is recommended to include corresponding descriptions to demonstrate that the product meets its intended performance.

 

Example from Annexes of IEC TR 60601-4-2

IEC 60601-4-2:2016 Annex D provides examples related to functions and performance indicators of ultrasound and imaging systems, emphasizing that the observed variations and deviations do not affect final diagnosis or treatment.

 

Similarly, for monitoring devices such as pulse oximeters and blood pressure monitors, the performance criteria can be defined according to clinical scenarios and risk analysis, considering the device design.

 

In addition to performance criteria for individual functions, IEC 60601-4-2 also includes examples for overall system immunity:

 

FDA IEC TR 60601-4-2 Standard Testing(图5)


> IEC 60601-4-2:2016 Annex D

 

For system-level equipment, the description of immunity criteria is no longer limited to single functions. Instead, it includes multiple stages such as startup, positioning, and final display, each with its corresponding criteria.

 

FDA IEC TR 60601-4-2 Standard Testing(图6)


These immunity criteria are crucial for testing laboratories during the evaluation process. Clear and detailed descriptions help laboratories quickly identify and resolve issues during testing, REDucing repeated communication and accelerating project progress.

 

Summary

IEC 60601-1-2 defines the EMC requirements related to medical device safety, but does not cover the performance of the device in various electromagnetic environments. In contrast, IEC 60601-4-2 addresses comprehensive performance issues under electromagnetic interference. While the two documents are related, their acceptance criteria differ significantly.

 

Compliance with IEC TR 60601-4-2 may be more challenging than compliance with IEC 60601-1-2.

 

Therefore, when working on 510(k) projects, there is no need to debate which standard applies — evidence of compliance with both IEC 60601-1-2 and IEC TR 60601-4-2 must be submitted to meet EMC requirements.

More: uk psti 5g compliance testing emi testing labs weee registration


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
What is Amazon TIC and How Can Sellers Achieve Com What is Amazon TIC and How Can Sellers Achieve Com
12 .16.2025
Amazon requires TIC-approved labs for Direct Validation. Tests must be full, under 1 year, ISO17025/...
2026 Battery UN38.3 Certification (Test Report) &a 2026 Battery UN38.3 Certification (Test Report) &a
12 .16.2025
2026 Battery UN38.3 Certification & MSDS: Update now to meet GHS Rev.11, IMDG 42-24 and IATA DGR...
What is the IEC 62680 Standard? Compliance Interpr What is the IEC 62680 Standard? Compliance Interpr
12 .16.2025
IEC 62680 sets unified USB-C interface & PD rules; EU now mandates compliance. JJR Lab provides ...
Amazon Japan December Compliance Requirements Amazon Japan December Compliance Requirements
12 .16.2025
Amazon Japan’s Dec-2025 rules require PSE, METI filing, TELEC, SAR (ARIB T66) and UN38.3. JJR provid...
How to Check a CPSC-Accepted Laboratory? How to Check a CPSC-Accepted Laboratory?
12 .16.2025
Verify CPSC-accepted labs via LabSearch to ensure CPSIA/CPC compliance for children’s products; meet...
WEEE Registration for Waste Electrical &Electr WEEE Registration for Waste Electrical &Electr
12 .15.2025
WEEE registration ensures EU recycling compliance per 2012/19/EU for EEE categories 1–6. Tests per E...
MSDS Chemical Safety Testing MSDS Chemical Safety Testing
12 .15.2025
JJR LAB provides MSDS chemical safety testing per legal standards, covering 16 sections on hazards, ...
What Are the Differences Between UK REACH and EU R What Are the Differences Between UK REACH and EU R
12 .15.2025
UK REACH, separate from EU REACH since 2021, lags in SVHC and Annex XVII updates; compliance testing...

Leave Your Message