EMC China Lab

What is the ASTM F2213-17 Testing Standard?

Views :
Update time : 2025-03-01

ASTM F2213 is a standard test method for measuring magnetically induced torque on medical devices in an MRI environment. The latest version is astm f2213-17.

 

China's JJR Laboratory provides testing and compliance services within the scope of ASTM F2213. This test method involves measuring the magnetically induced torque generated by the static magnetic field in an MRI environment and comparing this torque to the equivalent torque exerted by gravity on the implant.

 

What is the ASTM F2213-17 Testing Standard?(图1)


Acceptance Criteria

The acceptance criteria for this test shoULd be reasonable. If the maximum magnetically induced torque is less than the product of the longest dimension and weight of the medical device, then the magnetically induced torque is consideRED to be less than the worst-case torque due to gravity.

 

In such cases, it is assumed that the risk posed by the applied magnetically induced torque is no greater than the risks encountered during normal daily activities within Earth's gravitational field. This is considered a conservative approach.

 

Even larger torque values may not necessarily be hazardous. For example:

- The position of the device relative to adjacent tissue

- Tissue ingrowth

- Other mechanisms that may prevent device movement or generate forces from the magnetically induced torque that are less damaging than those caused by gravitational torque

 

This test method alone is not sufficient to determine whether an implant is safe in an MRI environment.

 

Torque Considerations

The magnetically induced torque considered in this standard is the static magnetic torque generated by the interaction between:

- The MRI’s static magnetic field

- The magnetization of the implant

 

This test method does not account for:

- DynaMIC torque generated by the interaction between the static field and eddy currents induced in rotating devices

- Torque induced by currents in conductors

 

ASTM F2213 Testing Scope

This test method involves measuring the magnetically induced torque on medical devices in an MRI environment and comparing this torque to user-specified acceptance criteria.

 

This test method does not cover:

- Magnetically induced translational force

- Tissue heating

- Device malfunction

- Imaging artifacts

- Acoustic noise

- Interference between devices

- Other potential safety concerns related to device function and MRI performance

 

Again, the torque considered here is the static magnetic torque generated by the interaction between:

- The MRI’s static magnetic field

- The magnetization of the implant

 

This test method does not include:

- Dynamic torque generated by the interaction between the static field and eddy currents in rotating devices

- Torque induced by currents in conductors

 

Significance and Use

This test is one of the methods used to determine whether a medical device may pose a risk in an MRI environment.

 

Other safety concerns include:

- Magnetically induced force

- RF heating

- Imaging artifacts

 

This standard is used together with:

- astm f2052 – Magnetically Induced Force Measurement

- astm f2182 – RF Heating Measurement

- ASTM F2119 – Imaging Artifact Evaluation

 

These standards collectively help evaluate medical device safety in an MRI environment.

 

Acceptance Criteria Summary

The maximum magnetically induced torque should be less than the product of the longest dimension and weight of the medical device.

 

This means the risk posed by the magnetically induced torque is considered no greater than the risks encountered during normal daily activities within Earth's gravitational field.

 

However, higher torque values may not necessarily be hazardous.

 

Measurement Units

Values expressed in SI units should be accepted as the standard.

Other measurement units are not included in this standard.


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
IEC 62471 Photobiological Safety of Lamps and Lamp IEC 62471 Photobiological Safety of Lamps and Lamp
02 .21.2026
IEC 62471 photobiological safety testing for LED lamps covers UV, blue light and IR risks, with risk...
New European Toy Standard EN 71-1:2026 New European Toy Standard EN 71-1:2026
02 .21.2026
EN 71-1:2026, the new European toy mechanical safety standard, updates testing requirements. JJR pro...
EN71 Series Standards Compliance February 13, 2026 EN71 Series Standards Compliance February 13, 2026
02 .21.2026
EN71 Compliance (Feb 13, 2026) — EU mandatory toy safety for children under 14. Updated EN71-2/4/5/7...
European Toy Safety Standard EN 71-20:2025 European Toy Safety Standard EN 71-20:2025
02 .20.2026
EN 71-20:2025 defines EU microbiological safety and preservation tests for toys with aqueous media; ...
EN 18031 Certification for Connected Devices on Am EN 18031 Certification for Connected Devices on Am
02 .20.2026
Amazon EU requires connected devices to meet EN 18031-1/2/3 cybersecurity standards; JJR lab provide...
Compliance Guide for Portable Batteries on Amazon Compliance Guide for Portable Batteries on Amazon
02 .20.2026
Amazon EU requires portable batteries to pass TIC audits by 15 Mar 2026 under EN/IEC 62368-1, IEC 62...
2026 EU SVHC Candidate List (253 Substances) 2026 EU SVHC Candidate List (253 Substances)
02 .15.2026
2026 EU SVHC Update: 253 substances confirmed by ECHA. REACH SVHC testing per EU standard; fees from...
LFGB Certification Cost and Timeline Guide LFGB Certification Cost and Timeline Guide
02 .15.2026
LFGB testing ensures food-contact safety under German standards. JJR Lab offers LFGB tests from $398...

Leave Your Message