EMC China Lab

What is the ASTM F2213-17 Testing Standard?

Views :
Update time : 2025-03-01

ASTM F2213 is a standard test method for measuring magnetically induced torque on medical devices in an MRI environment. The latest version is astm f2213-17.

 

China's JJR Laboratory provides testing and compliance services within the scope of ASTM F2213. This test method involves measuring the magnetically induced torque generated by the static magnetic field in an MRI environment and comparing this torque to the equivalent torque exerted by gravity on the implant.

 

What is the ASTM F2213-17 Testing Standard?(图1)


Acceptance Criteria

The acceptance criteria for this test should be reasonable. If the maximum magnetically induced torque is less than the product of the longest dimension and weight of the medical device, then the magnetically induced torque is considered to be less than the worst-case torque due to gravity.

 

In such cases, it is assumed that the risk posed by the applied magnetically induced torque is no greater than the risks encountered during normal daily activities within Earth's gravitational field. This is considered a conservative approach.

 

Even larger torque values may not necessarily be hazardous. For example:

- The position of the device relative to adjacent tissue

- Tissue ingrowth

- Other mechanisms that may prevent device movement or generate forces from the magnetically induced torque that are less damaging than those caused by gravitational torque

 

This test method alone is not sufficient to determine whether an implant is safe in an MRI environment.

 

Torque Considerations

The magnetically induced torque considered in this standard is the static magnetic torque generated by the interaction between:

- The MRI’s static magnetic field

- The magnetization of the implant

 

This test method does not account for:

- Dynamic torque generated by the interaction between the static field and eddy currents induced in rotating devices

- Torque induced by currents in conductors

 

ASTM F2213 Testing Scope

This test method involves measuring the magnetically induced torque on medical devices in an MRI environment and comparing this torque to user-specified acceptance criteria.

 

This test method does not cover:

- Magnetically induced translational force

- Tissue heating

- Device malfunction

- Imaging artifacts

- Acoustic noise

- Interference between devices

- Other potential safety concerns related to device function and MRI performance

 

Again, the torque considered here is the static magnetic torque generated by the interaction between:

- The MRI’s static magnetic field

- The magnetization of the implant

 

This test method does not include:

- Dynamic torque generated by the interaction between the static field and eddy currents in rotating devices

- Torque induced by currents in conductors

 

Significance and Use

This test is one of the methods used to determine whether a medical device may pose a risk in an MRI environment.

 

Other safety concerns include:

- Magnetically induced force

- RF heating

- Imaging artifacts

 

This standard is used together with:

- astm f2052 – Magnetically Induced Force Measurement

- astm f2182 – RF Heating Measurement

- ASTM F2119 – Imaging Artifact Evaluation

 

These standards collectively help evaluate medical device safety in an MRI environment.

 

Acceptance Criteria Summary

The maximum magnetically induced torque should be less than the product of the longest dimension and weight of the medical device.

 

This means the risk posed by the magnetically induced torque is considered no greater than the risks encountered during normal daily activities within Earth's gravitational field.

 

However, higher torque values may not necessarily be hazardous.

 

Measurement Units

Values expressed in SI units should be accepted as the standard.

Other measurement units are not included in this standard.


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
What is the EN 60598-1:2021 Test Standard? What is the EN 60598-1:2021 Test Standard?
08 .22.2026
Discover the EN 60598-1:2021 test standard for luminaire safety and performance. Trust JJR Lab for e...
What is the EN IEC 60335-1 Standard? What is the EN IEC 60335-1 Standard?
08 .22.2026
What is the EN IEC 60335-1 testing standard? It is the core safety requirement for EU home appliance...
EU Authorized Representative Medical Device EU Authorized Representative Medical Device
08 .21.2026
Need an EU Authorized Representative for medical devices? JJR Lab provides expert CE MDR compliance ...
Electromagnetic Compatibility Testing Services Electromagnetic Compatibility Testing Services
08 .21.2026
JJR Laboratory provides accredited electromagnetic compatibility (EMC) testing services. Ensure your...
Faucets and Showerheads Must Upload DOE and CEC Nu Faucets and Showerheads Must Upload DOE and CEC Nu
08 .21.2026
Amazon US requires mandatory DOE & CEC certification for faucets and showerheads. Meet flow rate...
Showerhead DOE Energy Efficiency Application Showerhead DOE Energy Efficiency Application
08 .21.2026
Get US DOE certification for showerheads with JJR Lab. We explain test standards, 9.5L/min limit at ...
How to Get Shower Head DOE Certification How to Get Shower Head DOE Certification
08 .20.2026
Secure your shower head DOE certification for Amazon US with JJR LAB. We ensure full compliance with...
Amazon US DOE Certification and CCMS for Showerhea Amazon US DOE Certification and CCMS for Showerhea
08 .20.2026
Amazon showerhead DOE & CEC compliance: JJR Lab provides 10 CFR 430 and ASME A112.18.1 testing p...

Leave Your Message