Main Reference Standards

◆ CISPR 11/EN 55011/GB 4824:

   "Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment - Electromagnetic Disturbance Characteristics - Limits and Methods of Measurement"

◆ iec 60601-1-2/YY 0505:

   "Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests"

◆ iec 61326-1/GB T18268.1:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 1: General Requirements"

◆ IEC 61326-2-6/GB T18268.26:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 26: Specific Requirements - In Vitro Diagnostic (IVD) Equipment"

◆ iec 61000-3-2/GB 17625.1:

   "Electromagnetic Compatibility - Limits - Limits for Harmonic Current Emissions (Equipment Input Current ≤16A per Phase)"

◆ IEC 61000-3-3/GB 17625.2:

   "Electromagnetic Compatibility - Limits - Limitation of Voltage Changes, Voltage Fluctuations, and Flicker in Public Low-Voltage Supply Systems, for Equipment with Rated Current ≤16A and Subject to Conditional Connection"

 

emi test Items

◆ Conducted Disturbance Voltage, 150kHz-30MHz, Quasi-Peak/Average

◆ Radiated Disturbance (Magnetic Field), 9kHz-30MHz, Quasi-Peak

◆ Radiated Disturbance (Electric Field), 150kHz-1GHz, Quasi-Peak; 1GHz-18GHz, Peak/Weighted

◆ Harmonic Current Emission Test, According to IEC 61000-3-2

◆ Voltage Fluctuation and Flicker Test, According to IEC 61000-3-3

 

EMS Test Items

◆ Electrostatic Discharge Immunity, Contact Discharge 6kV, Air Discharge 8kV

◆ Radiated Radio-Frequency Electromagnetic Field Immunity, Maximum Test Field Strength 10V/m

◆ Electrical Fast Transient/Burst Immunity, Maximum Test Level 2kV

◆ Surge Immunity, Maximum Test Level 2kV

◆ Radio-Frequency Conducted Immunity, Maximum Test Level 10V

◆ Power Frequency Magnetic Field Immunity, Maximum Test Field Strength 3A/m

◆ Voltage Dips, Short Interruptions, and Voltage Variations Immunity, According to IEC 61000-4-11


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
UKCA Certification Guide UKCA Certification Guide
11 .12.2026
Master 2027 UKCA certification with JJR LAB‘s expert guide. As an ISO/IEC 17025, CMA, and CNAS accre...
2027 Amazon Requires UL2743 Test Report for Emerge 2027 Amazon Requires UL2743 Test Report for Emerge
11 .12.2026
Need an Amazon UL2743 test report for car jump starters? JJR LAB provides ISO/IEC 17025 accredited t...
2027 Compliance Guide for Lighting Fixtures 2027 Compliance Guide for Lighting Fixtures
11 .12.2026
Ensure 2027 lighting fixture compliance with JJR Laboratory. As an ISO/IEC 17025, CMA and CNAS accre...
2027 EU Export CE Certification Guidelines 2027 EU Export CE Certification Guidelines
11 .12.2026
Secure 2027 EU export compliance with JJR LAB! We provide expert CE & CE RED certification testi...
Cybersecurity EN 18031 Certifications Cybersecurity EN 18031 Certifications
11 .12.2026
Ensure EU RED compliance with EN 18031 cybersecurity testing for smart products. JJR LAB provides IS...
What Certifications Are Required for Japan Exports What Certifications Are Required for Japan Exports
11 .12.2026
Exporting to Japan? JJR Lab provides ISO/IEC 17025, CMA & CNAS accredited testing for mandatory ...
What is 16 CFR 1260 Compliance for Window Covering What is 16 CFR 1260 Compliance for Window Covering
11 .11.2026
Get 16 CFR 1260 compliance testing for custom window coverings at JJR Laboratory. ISO/IEC 17025 accr...
Differences Between F2923 and F2999 Differences Between F2923 and F2999
11 .11.2026
Need ASTM F2923 & F2999 testing for US Amazon jewelry compliance? JJR LAB is a CMA & CNAS ac...

Leave Your Message