EMC China Lab

Medical Device EMC Test Items & Standards

Views :
Update time : 2024-04-26

Main Reference Standards

◆ CISPR 11/EN 55011/GB 4824:

   "Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment - Electromagnetic Disturbance Characteristics - Limits and Methods of Measurement"

iec 60601-1-2/YY 0505:

   "Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests"

iec 61326-1/GB T18268.1:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 1: General Requirements"

◆ IEC 61326-2-6/GB T18268.26:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 26: Specific Requirements - In Vitro Diagnostic (IVD) Equipment"

◆ IEC 61000-3-2/GB 17625.1:

   "Electromagnetic Compatibility - Limits - Limits for Harmonic Current Emissions (Equipment Input Current ≤16A per Phase)"

◆ IEC 61000-3-3/GB 17625.2:

   "Electromagnetic Compatibility - Limits - Limitation of Voltage Changes, Voltage Fluctuations, and Flicker in Public Low-Voltage Supply Systems, for Equipment with Rated Current ≤16A and Subject to Conditional Connection"

 

emi test Items

◆ Conducted Disturbance Voltage, 150kHz-30MHz, Quasi-Peak/Average

◆ Radiated Disturbance (Magnetic Field), 9kHz-30MHz, Quasi-Peak

◆ Radiated Disturbance (Electric Field), 150kHz-1GHz, Quasi-Peak; 1GHz-18GHz, Peak/Weighted

◆ Harmonic Current Emission Test, According to IEC 61000-3-2

◆ Voltage Fluctuation and Flicker Test, According to IEC 61000-3-3

 

EMS Test Items

◆ Electrostatic Discharge Immunity, Contact Discharge 6kV, Air Discharge 8kV

◆ Radiated Radio-Frequency Electromagnetic Field Immunity, Maximum Test Field Strength 10V/m

◆ Electrical Fast Transient/Burst Immunity, Maximum Test Level 2kV

◆ Surge Immunity, Maximum Test Level 2kV

◆ Radio-Frequency Conducted Immunity, Maximum Test Level 10V

◆ Power Frequency Magnetic Field Immunity, Maximum Test Field Strength 3A/m

◆ Voltage Dips, Short Interruptions, and Voltage Variations Immunity, According to IEC 61000-4-11


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
Infant Support Pillow 16 CFR 1243/1242 & ASTM Infant Support Pillow 16 CFR 1243/1242 & ASTM
12 .19.2025
Infant Support & Nursing Pillows must meet CPSC 2025 rules, including 16 CFR 1243/1242 & AST...
BRM Registration Card Under CFR Part 1130 Regulati BRM Registration Card Under CFR Part 1130 Regulati
12 .19.2025
Required for CFR Part 1130 infant products, the BRM Card supports CPC compliance and U.S. address ve...
How to get a D-U-N-S® Number for US FDA Registrati How to get a D-U-N-S® Number for US FDA Registrati
12 .18.2025
Get your D-U-N-S® Number for FDA registration. JJR LAB helps medical device firms apply fast, meet t...
Household Massage Devices Compliance in the China Household Massage Devices Compliance in the China
12 .18.2025
JJR LAB ensures household massage devices comply in China & Japan: CQC (GB4706.1/10), PSE (J/IEC...
Compliance for the Global In Vitro Diagnostic (IVD Compliance for the Global In Vitro Diagnostic (IVD
12 .18.2025
JJR LAB, ISO/IEC 17025 certified, offers one-stop IVD device compliance testing & certification ...
Compliance Guide for Nebulizers in European and Am Compliance Guide for Nebulizers in European and Am
12 .18.2025
JJR LAB provides CE/FDA nebulizer testing (IEC 60601, ISO 27427, ISO 18562, ISO 10993) covering perf...
Cybersecurity Certification Service for EU RED Dir Cybersecurity Certification Service for EU RED Dir
12 .18.2025
JJR Lab provides RED 3.3(d/e/f) compliance via EN18031-1/2/3 testing, offering asset review, securit...
ANATEL Certification Compliance Guide for Brazil M ANATEL Certification Compliance Guide for Brazil M
12 .18.2025
ANATEL compliance ensures RF/EMC/safety testing to Brazil standards; required for wireless/network d...

Leave Your Message