EMC China Lab

Medical Device EMC Test Items & Standards

Views :
Update time : 2024-04-26

Main Reference Standards

◆ CISPR 11/EN 55011/GB 4824:

   "Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment - Electromagnetic Disturbance Characteristics - Limits and Methods of Measurement"

◆ iec 60601-1-2/YY 0505:

   "Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests"

◆ iec 61326-1/GB T18268.1:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 1: General Requirements"

◆ IEC 61326-2-6/GB T18268.26:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 26: Specific Requirements - In Vitro Diagnostic (IVD) Equipment"

◆ iec 61000-3-2/GB 17625.1:

   "Electromagnetic Compatibility - Limits - Limits for Harmonic Current Emissions (Equipment Input Current ≤16A per Phase)"

◆ IEC 61000-3-3/GB 17625.2:

   "Electromagnetic Compatibility - Limits - Limitation of Voltage Changes, Voltage Fluctuations, and Flicker in Public Low-Voltage Supply Systems, for Equipment with Rated Current ≤16A and Subject to Conditional Connection"

 

emi test Items

◆ Conducted Disturbance Voltage, 150kHz-30MHz, Quasi-Peak/Average

◆ Radiated Disturbance (Magnetic Field), 9kHz-30MHz, Quasi-Peak

◆ Radiated Disturbance (Electric Field), 150kHz-1GHz, Quasi-Peak; 1GHz-18GHz, Peak/Weighted

◆ Harmonic Current Emission Test, According to IEC 61000-3-2

◆ Voltage Fluctuation and Flicker Test, According to IEC 61000-3-3

 

EMS Test Items

◆ Electrostatic Discharge Immunity, Contact Discharge 6kV, Air Discharge 8kV

◆ Radiated Radio-Frequency Electromagnetic Field Immunity, Maximum Test Field Strength 10V/m

◆ Electrical Fast Transient/Burst Immunity, Maximum Test Level 2kV

◆ Surge Immunity, Maximum Test Level 2kV

◆ Radio-Frequency Conducted Immunity, Maximum Test Level 10V

◆ Power Frequency Magnetic Field Immunity, Maximum Test Field Strength 3A/m

◆ Voltage Dips, Short Interruptions, and Voltage Variations Immunity, According to IEC 61000-4-11


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
What is Cytotoxicity ISO 10993-5 What is Cytotoxicity ISO 10993-5
09 .29.2026
ISO 10993-5 cytotoxicity testing at JJR LAB uses cell culture hardware and analyzers to measure cell...
What is Hemocompatibility Testing ISO 10993? What is Hemocompatibility Testing ISO 10993?
09 .29.2026
ISO 10993-4 hemocompatibility testing uses blood-contact test systems, flow probes and analyzers to ...
What is ISO 10993-5 Tests for In Vitro Cytotoxicit What is ISO 10993-5 Tests for In Vitro Cytotoxicit
09 .29.2026
ISO 10993-5 cytotoxicity testing at JJR LAB uses L929 cells, MTT hardware and extraction methods to ...
Electronic Product IEC 62368-1 Safety Testing Electronic Product IEC 62368-1 Safety Testing
09 .28.2026
JJR Lab provides IEC 62368-1 safety testing for AV and ICT products, covering HBSE, electric shock, ...
What is an EU SDS (Safety Data Sheet)? What is an EU SDS (Safety Data Sheet)?
09 .25.2026
JJR Lab explains EU SDS requirements under REACH, CLP and EU 2020/878, including 16-section Safety D...
What is the CA Prop 65 Warning Label? What is the CA Prop 65 Warning Label?
09 .24.2026
JJR Lab explains CA Prop 65 warning labels and compliance testing for lead, cadmium, phthalates and ...
2026 Amazon TIC Mandatory Audit Catalog 2026 Amazon TIC Mandatory Audit Catalog
09 .24.2026
2026 Amazon TIC Mandatory Audit Catalog by JJR Lab covers UN38.3, UL 4200A, ASTM F963, FCC Part 15C,...
Australian Electrical EESS and RCM Compliance Australian Electrical EESS and RCM Compliance
09 .24.2026
Australian EESS and RCM compliance with AS/NZS 4417, AS/NZS 3199 and IEC 60974-1 testing. JJR Testin...

Leave Your Message