EMC China Lab

Medical Device EMC Test Items & Standards

Views :
Update time : 2024-04-26

Main Reference Standards

◆ CISPR 11/EN 55011/GB 4824:

   "Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment - Electromagnetic Disturbance Characteristics - Limits and Methods of Measurement"

iec 60601-1-2/YY 0505:

   "Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests"

iec 61326-1/GB T18268.1:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 1: General Requirements"

◆ IEC 61326-2-6/GB T18268.26:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 26: Specific Requirements - In Vitro Diagnostic (IVD) Equipment"

iec 61000-3-2/GB 17625.1:

   "Electromagnetic Compatibility - Limits - Limits for Harmonic Current Emissions (Equipment Input Current ≤16A per Phase)"

◆ IEC 61000-3-3/GB 17625.2:

   "Electromagnetic Compatibility - Limits - Limitation of Voltage Changes, Voltage Fluctuations, and Flicker in Public Low-Voltage Supply Systems, for Equipment with Rated Current ≤16A and Subject to Conditional Connection"

 

emi test Items

◆ Conducted Disturbance Voltage, 150kHz-30MHz, Quasi-Peak/Average

◆ Radiated Disturbance (Magnetic Field), 9kHz-30MHz, Quasi-Peak

◆ Radiated Disturbance (Electric Field), 150kHz-1GHz, Quasi-Peak; 1GHz-18GHz, Peak/Weighted

◆ Harmonic Current Emission Test, According to IEC 61000-3-2

◆ Voltage Fluctuation and Flicker Test, According to IEC 61000-3-3

 

EMS Test Items

◆ Electrostatic Discharge Immunity, Contact Discharge 6kV, Air Discharge 8kV

◆ Radiated Radio-Frequency Electromagnetic Field Immunity, Maximum Test Field Strength 10V/m

◆ Electrical Fast Transient/Burst Immunity, Maximum Test Level 2kV

◆ Surge Immunity, Maximum Test Level 2kV

◆ Radio-Frequency Conducted Immunity, Maximum Test Level 10V

◆ Power Frequency Magnetic Field Immunity, Maximum Test Field Strength 3A/m

◆ Voltage Dips, Short Interruptions, and Voltage Variations Immunity, According to IEC 61000-4-11


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
California 65 Testing Service Consultation California 65 Testing Service Consultation
09 .17.2026
Choose JJR Lab for reliable California 65 testing services. We ensure your products strictly comply ...
REACH Chemical Testing Services REACH Chemical Testing Services
09 .17.2026
Ensure EU compliance with JJR Lab‘s REACH chemical testing services. We deliver expert SVHC testing ...
Electronic Vaporizers EU TPD / US PMTA / UK MHRA Electronic Vaporizers EU TPD / US PMTA / UK MHRA
09 .17.2026
JJR Lab provides expert electronic vaporizer testing to global standards like GB 41700, ensuring EU ...
ISO 10993 Biocompatibility Standards ISO 10993 Biocompatibility Standards
09 .16.2026
Ensure medical device safety and global compliance with JJR Lab. We provide expert ISO 10993 biocomp...
Medical TPV ISO 10993 Biocompatibility Testing Medical TPV ISO 10993 Biocompatibility Testing
09 .16.2026
Ensure Medical TPV compliance via ISO 10993 biocompatibility testing at JJR Laboratory. We offer che...
FDA Requirements for Biocompatibility under ISO 10 FDA Requirements for Biocompatibility under ISO 10
09 .15.2026
Ensure your medical devices meet FDA biocompatibility requirements. JJR Laboratory offers expert ISO...
Amazon Spain WEEE Compliance Deadline: September 3 Amazon Spain WEEE Compliance Deadline: September 3
09 .15.2026
Amazon Spain WEEE Compliance Deadline: September 30, 2026. Register your electrical equipment with J...
What is the ASTM F2264 Certification Test? What is the ASTM F2264 Certification Test?
09 .14.2026
What is the ASTM F2264 certification test? It is the US safety standard for non-powered scooters. Ch...

Leave Your Message