EMC China Lab

Medical Device EMC Test Items & Standards

Views :
Update time : 2024-04-26

Main Reference Standards

◆ CISPR 11/EN 55011/GB 4824:

   "Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment - Electromagnetic Disturbance Characteristics - Limits and Methods of Measurement"

◆ IEC 60601-1-2/YY 0505:

   "Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests"

◆ IEC 61326-1/GB T18268.1:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 1: General Requirements"

◆ IEC 61326-2-6/GB T18268.26:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 26: Specific Requirements - In Vitro Diagnostic (IVD) Equipment"

◆ IEC 61000-3-2/GB 17625.1:

   "Electromagnetic Compatibility - Limits - Limits for Harmonic Current Emissions (Equipment Input Current ≤16A per Phase)"

◆ IEC 61000-3-3/GB 17625.2:

   "Electromagnetic Compatibility - Limits - Limitation of Voltage Changes, Voltage Fluctuations, and Flicker in Public Low-Voltage Supply Systems, for Equipment with Rated Current ≤16A and Subject to Conditional Connection"

 

EMI Test Items

◆ Conducted Disturbance Voltage, 150kHz-30MHz, Quasi-Peak/Average

◆ Radiated Disturbance (Magnetic Field), 9kHz-30MHz, Quasi-Peak

◆ Radiated Disturbance (Electric Field), 150kHz-1GHz, Quasi-Peak; 1GHz-18GHz, Peak/Weighted

◆ Harmonic Current Emission Test, According to IEC 61000-3-2

◆ Voltage Fluctuation and Flicker Test, According to IEC 61000-3-3

 

EMS Test Items

◆ Electrostatic Discharge Immunity, Contact Discharge 6kV, Air Discharge 8kV

◆ Radiated Radio-Frequency Electromagnetic Field Immunity, Maximum Test Field Strength 10V/m

◆ Electrical Fast Transient/Burst Immunity, Maximum Test Level 2kV

◆ Surge Immunity, Maximum Test Level 2kV

◆ Radio-Frequency Conducted Immunity, Maximum Test Level 10V

◆ Power Frequency Magnetic Field Immunity, Maximum Test Field Strength 3A/m

◆ Voltage Dips, Short Interruptions, and Voltage Variations Immunity, According to IEC 61000-4-11


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
FCC ID Certification Compliance Certificate FCC ID Certification Compliance Certificate
04 .30.2025
FCC ID certifies RF products meet U.S. standards for safety and interference. JJR Lab provides testi...
How many types of FCC certifications are there? How many types of FCC certifications are there?
04 .30.2025
FCC certification includes SDoC and FCC ID for electronic and wireless devices. JJR Lab provides tes...
Australia RCM LOGO Certification Australia RCM LOGO Certification
04 .30.2025
RCM certification combines safety & EMC for AU/NZ market access. JJR Lab provides testing to hel...
ANATEL Certification Process in Brazil ANATEL Certification Process in Brazil
04 .30.2025
ANATEL certification is required for telecom products entering Brazil. JJR Lab provides testing to e...
Faucet European Standard EN 817 Testing Faucet European Standard EN 817 Testing
04 .30.2025
JJR Lab provides faucet testing and certification according to European standards EN 817, EN 200, EN...
ISO 17025 Laboratory Test Report ISO 17025 Laboratory Test Report
04 .30.2025
Amazon product removals may occur due to missing ILAC ISO17025 reports or label issues. JJR Lab prov...
Temu Requires UN 38.3 Certification Temu Requires UN 38.3 Certification
04 .30.2025
TEMU requires UN 38.3 certification for battery transport. JJR Testing Laboratory offers professiona...
What is the Amazon ASTM F2057 Test Report? What is the Amazon ASTM F2057 Test Report?
04 .30.2025
The ASTM F2057-23 standard ensures safety for clothing storage units, focusing on stability, flammab...

Leave Your Message