EMC China Lab

Medical Device EMC Test Items & Standards

Views :
Update time : 2024-04-26

Main Reference Standards

◆ CISPR 11/EN 55011/GB 4824:

   "Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment - Electromagnetic Disturbance Characteristics - Limits and Methods of Measurement"

iec 60601-1-2/YY 0505:

   "Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests"

iec 61326-1/GB T18268.1:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 1: General Requirements"

◆ IEC 61326-2-6/GB T18268.26:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 26: Specific Requirements - In Vitro Diagnostic (IVD) Equipment"

◆ IEC 61000-3-2/GB 17625.1:

   "Electromagnetic Compatibility - Limits - Limits for Harmonic Current Emissions (Equipment Input Current ≤16A per Phase)"

◆ IEC 61000-3-3/GB 17625.2:

   "Electromagnetic Compatibility - Limits - Limitation of Voltage Changes, Voltage Fluctuations, and Flicker in Public Low-Voltage Supply Systems, for Equipment with Rated Current ≤16A and Subject to Conditional Connection"

 

emi test Items

◆ Conducted Disturbance Voltage, 150kHz-30MHz, Quasi-Peak/Average

◆ Radiated Disturbance (Magnetic Field), 9kHz-30MHz, Quasi-Peak

◆ Radiated Disturbance (Electric Field), 150kHz-1GHz, Quasi-Peak; 1GHz-18GHz, Peak/Weighted

◆ Harmonic Current Emission Test, According to IEC 61000-3-2

◆ Voltage Fluctuation and Flicker Test, According to IEC 61000-3-3

 

EMS Test Items

◆ Electrostatic Discharge Immunity, Contact Discharge 6kV, Air Discharge 8kV

◆ Radiated Radio-Frequency Electromagnetic Field Immunity, Maximum Test Field Strength 10V/m

◆ Electrical Fast Transient/Burst Immunity, Maximum Test Level 2kV

◆ Surge Immunity, Maximum Test Level 2kV

◆ Radio-Frequency Conducted Immunity, Maximum Test Level 10V

◆ Power Frequency Magnetic Field Immunity, Maximum Test Field Strength 3A/m

◆ Voltage Dips, Short Interruptions, and Voltage Variations Immunity, According to IEC 61000-4-11


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
Mobile Batteries for Garments: Japan Compliance | Mobile Batteries for Garments: Japan Compliance |
01 .14.2026
Mobile batteries for garments must meet Japan PSE safety rules; JJR Lab tests for fire, smoke, overh...
(FAQs) about Japan’s PSE Four Product Safety Laws (FAQs) about Japan’s PSE Four Product Safety Laws
01 .14.2026
From Dec 25, 2025, overseas sellers of Japan’s Four Product Safety Law items must register, label wi...
FCC 47 CFR Part 2.1093 Radio Frequency Exposure Te FCC 47 CFR Part 2.1093 Radio Frequency Exposure Te
01 .14.2026
JJR Lab performs FCC 47 CFR Part 2.1093 RF exposure tests (SAR, PD, EMF) on phones, tablets, laptops...
Compliance Testing for EU Food Contact Materials R Compliance Testing for EU Food Contact Materials R
01 .14.2026
JJR Lab offers EU Food Contact Materials compliance testing per Regulation 1935/2004, including over...
Compliance for Portable Power Supplies on Amazon U Compliance for Portable Power Supplies on Amazon U
01 .14.2026
Ensure Amazon US/Canada portable power supply compliance: test all ASINs via JJR Lab per UL 2056/UL ...
Compliance for Portable Power Stations on Amazon E Compliance for Portable Power Stations on Amazon E
01 .14.2026
AmazonEU portable power compliance requires IEC 62368-1, IEC 62133-2 and UN38.3 testing by JJR Lab b...
Compliance for Baby Feeding Products on Amazon Eur Compliance for Baby Feeding Products on Amazon Eur
01 .14.2026
Compliance for Baby Feeding Products on Amazon Europe requires baby bottles meet EN14350 and EU10/20...
Compliance Guide for Bed Rails on Amazon Australia Compliance Guide for Bed Rails on Amazon Australia
01 .14.2026
Compliance Guide for Bed Rails on Amazon AU: meet ASTM F2085-19 or BS7972, list model & brand, s...

Leave Your Message