Main Reference Standards

◆ CISPR 11/EN 55011/GB 4824:

   "Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment - Electromagnetic Disturbance Characteristics - Limits and Methods of Measurement"

◆ IEC 60601-1-2/YY 0505:

   "Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests"

◆ IEC 61326-1/GB T18268.1:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 1: General Requirements"

◆ IEC 61326-2-6/GB T18268.26:

   "Electrical Equipment for Measurement, Control, and Laboratory Use - EMC Requirements - Part 26: Specific Requirements - In Vitro Diagnostic (IVD) Equipment"

◆ IEC 61000-3-2/GB 17625.1:

   "Electromagnetic Compatibility - Limits - Limits for Harmonic Current Emissions (Equipment Input Current ≤16A per Phase)"

◆ IEC 61000-3-3/GB 17625.2:

   "Electromagnetic Compatibility - Limits - Limitation of Voltage Changes, Voltage Fluctuations, and Flicker in Public Low-Voltage Supply Systems, for Equipment with Rated Current ≤16A and Subject to Conditional Connection"

 

EMI Test Items

◆ Conducted Disturbance Voltage, 150kHz-30MHz, Quasi-Peak/Average

◆ Radiated Disturbance (Magnetic Field), 9kHz-30MHz, Quasi-Peak

◆ Radiated Disturbance (Electric Field), 150kHz-1GHz, Quasi-Peak; 1GHz-18GHz, Peak/Weighted

◆ Harmonic Current Emission Test, According to IEC 61000-3-2

◆ Voltage Fluctuation and Flicker Test, According to IEC 61000-3-3

 

EMS Test Items

◆ Electrostatic Discharge Immunity, Contact Discharge 6kV, Air Discharge 8kV

◆ Radiated Radio-Frequency Electromagnetic Field Immunity, Maximum Test Field Strength 10V/m

◆ Electrical Fast Transient/Burst Immunity, Maximum Test Level 2kV

◆ Surge Immunity, Maximum Test Level 2kV

◆ Radio-Frequency Conducted Immunity, Maximum Test Level 10V

◆ Power Frequency Magnetic Field Immunity, Maximum Test Field Strength 3A/m

◆ Voltage Dips, Short Interruptions, and Voltage Variations Immunity, According to IEC 61000-4-11


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
Amazon UL Standard Test Report Amazon UL Standard Test Report
04 .26.2025
China JJR Lab provides UL testing for Amazon listings, ensuring ISO 17025 compliance for powered pro...
When Can FCC ID Modifications Be Filed? When Can FCC ID Modifications Be Filed?
04 .26.2025
FCC ID modification includes Class I, II, III, and Change ID. JJR Lab China provides reliable, effic...
LoRa Certification Testing Laboratory LoRa Certification Testing Laboratory
04 .26.2025
LoRa Certification Testing by China JJR Lab: We offer full testing for LoRa modules and gateways acr...
Blood Pressure Monitor Certification Testing Servi Blood Pressure Monitor Certification Testing Servi
04 .26.2025
Blood pressure monitor testing per global and China standards, including EMC, safety, performance, a...
ECG Device Certification Testing ECG Device Certification Testing
04 .26.2025
ECG device certification testing, including EMC, safety, performance, biocompatibility, and steriliz...
Pulse Oximeter Certification and Testing Standards Pulse Oximeter Certification and Testing Standards
04 .26.2025
Pulse oximeters measure SpO₂ using light technology. JJR Laboratory in China provides certification ...
IVD Medical Device GB 4793:2024 Test Report IVD Medical Device GB 4793:2024 Test Report
04 .26.2025
GB 4793:2024 IVD safety testing now mandatory—JJR Laboratory China offers certified testing services...
IECEE CBTL Testing Laboratory for IVD Medical Devi IECEE CBTL Testing Laboratory for IVD Medical Devi
04 .26.2025
China JJR Laboratory, accredited by IECEE CBTL, offers testing for MED & IVD equipment, providin...

Leave Your Message