EMC China Lab

China Mask Certification Testing Laboratory

Views :
Update time : 2025-03-28

In response to market and consumer concerns about mask quality and safety, as well as the demand for mask exports overseas, China JJR Laboratory has invested over one million USD in advanced mask testing equipment and actively sought recognition from national and international organizations to provide fast, flexible, and professional internationally recognized mask testing services. Today, JJR has obtained dual recognition from the China National Accreditation Service for Conformity Assessment (CNAS) and the International Accreditation Service (IAS), offering mask testing services accredited by the International Laboratory Accreditation Cooperation (ILAC).

 

China JJR Laboratory provides comprehensive international standard testing such as: Australia AS4381: 2015, China YY0469-2011, YY/T0969-2013, and GB15979, European Union EN 14683: 2019+AC: 2019, and the United States ASTM F2100-19, meeting the requirements for mask exports to different countries and registration.

 

ASTM F2100-19 Testing Methods:

- Bacterial Filtration Efficiency (BFE) %:

- Level 1: ≥95

- Level 2: ≥98

- Level 3: ≥98

- Testing Method: ASTM F2101

 

- Particulate Filtration Efficiency (PFE) % @0.1 micron:

- Level 1: ≥95

- Level 2: ≥98

- Level 3: ≥98

- Testing Method: ASTM F2299

 

- Pressure Differential (mmH₂O/cm²):

- Level 1: <5.0

- Level 2: <6.0

- Level 3: <6.0

- Testing Method: EN 14683

 

- Synthetic Blood Penetration Resistance (mmHg):

- Level 1: 80

- Level 2: 120

- Level 3: 160

- Testing Method: ASTM F1862

 

- Flammability Test:

- Level 1: Class 1

- Level 2: Class 1

- Level 3: Class 1

- Testing Method: 16 CFR 1610

 

EN 14683: 2019 Testing Methods:

- Bacterial Filtration Efficiency (BFE) %:

- Level I: ≥95

- Level II: ≥98

- Level IIR: ≥98

- Testing Method: EN 14683

 

- Pressure Differential (Delta P) (Pa/cm²):

- Level I: <40

- Level II: <40

- Level IIR: <60

- Testing Method: EN 14683

 

- Splash Resistance (kPa):

- Level I: Not Required

- Level II: Not Required

- Level IIR: ≥16.0 (120 mmHg)

- Testing Method: EN 14683

 

- Microbial Cleanliness (CFU/g):

- Level I: ≤30

- Level II: ≤30

- Level IIR: ≤30

- Testing Method: EN 14683

 

AS 4381: 2015 Testing Methods:

- Bacterial Filtration Efficiency (BFE) %:

- Level 1: ≥95

- Level 2: ≥98

- Level 3: ≥98

- Testing Method: EN 14683

 

China Mask Certification Testing Laboratory(图1)


- Pressure Differential (mmH₂O/cm²):

- Level 1: <4.0

- Level 2: <5.0

- Level 3: <5.0

- Testing Method: EN 14683

 

- Synthetic Blood Penetration Resistance (mmHg):

- Level 1: 80

- Level 2: 120

- Level 3: 160

- Testing Method: ISO 22609

 

biocompatibility iso 10993 testing Methods:

- Cell Toxicity: iso 10993-5

- Irritation: iso 10993-10

- Skin Sensitization: ISO 10993-10

 

In addition to the above standards, our center also offers various mask tests to meet different market requirements, including:

 

- Viral Filtration Efficiency (VFE) %

- GB 15979 Microbial Indicator Requirements

- Antibacterial Efficiency Testing

- Antiviral Efficiency Testing

 

Mask Testing Equipment:

To meet the large demand for mask testing and to ensure timely reporting, China JJR Laboratory has invested over one million USD in advanced mask testing equipment to immediately improve the mask testing capacity and speed.

 


Email:hello@jjrlab.com


Leave Your Message


Write your message here and send it to us


Related News
Read More >>
ISO 10993 Biocompatibility Standards ISO 10993 Biocompatibility Standards
09 .16.2026
Ensure medical device safety and global compliance with JJR Lab. We provide expert ISO 10993 biocomp...
Medical TPV ISO 10993 Biocompatibility Testing Medical TPV ISO 10993 Biocompatibility Testing
09 .16.2026
Ensure Medical TPV compliance via ISO 10993 biocompatibility testing at JJR Laboratory. We offer che...
FDA Requirements for Biocompatibility under ISO 10 FDA Requirements for Biocompatibility under ISO 10
09 .15.2026
Ensure your medical devices meet FDA biocompatibility requirements. JJR Laboratory offers expert ISO...
Amazon Spain WEEE Compliance Deadline: September 3 Amazon Spain WEEE Compliance Deadline: September 3
09 .15.2026
Amazon Spain WEEE Compliance Deadline: September 30, 2026. Register your electrical equipment with J...
What is the ASTM F2264 Certification Test? What is the ASTM F2264 Certification Test?
09 .14.2026
What is the ASTM F2264 certification test? It is the US safety standard for non-powered scooters. Ch...
What is ASTM F2194 Certification Testing? What is ASTM F2194 Certification Testing?
09 .14.2026
Learn how to get ASTM F2194 certification for infant bassinets. JJR Lab offers CPSC-accepted safety ...
What is the ASTM F2167 Certification Test? What is the ASTM F2167 Certification Test?
09 .11.2026
Pass Amazon compliance with ASTM F2167 infant bouncer seat testing at JJR Laboratory. We offer exper...
What is the ASTM F2050 Certification Test? What is the ASTM F2050 Certification Test?
09 .11.2026
Need ASTM F2050 certification for hand-held infant carriers? JJR Laboratory provides professional te...

Leave Your Message